A Technique to Evaluate MRI-induced Electric Fields at the Ends of Practical Implanted Lead

来自NERCN
Admin讨论 | 贡献2024年1月22日 (一) 13:59的版本 (Edit summary)
(差异) ←上一版本 | 最后版本 (差异) | 下一版本→ (差异)

推荐理由

基于互易定理与惠更斯原理提出一种有效评估植入物金属导线局部感应电场或感应电压新技术,解耦微观尺度金属导线模拟/测量与宏观人体模拟,简化导线MRI感应加热和植入物感应电压估计过程。

文章简介
期刊 IEEE Transactions on Microwave Theory and Techniques
发表年份 2014
DOI 10.1109/TMTT.2014.2376523
类型 研究性工作
领域 磁共振兼容
引用量 134
推荐信息
推荐人 张肇源
审核 姜老师
推荐小组 磁共振小组

摘要

This paper presents a novel technique for efficient evaluation of magnetic resonance imaging (MRI)-induced electric fields or induced voltages in the vicinity of implanted metallic leads. The technique is based on the reciprocity theorem in conjunction with the Huygens Principle. This approach allows one to decouple the micro-scale metallic lead simulation/measurement from the macro-level phantom human simulations within the MRI scanners. Consequently, the estimation of MRI-induced heating on an implanted lead, and the induced voltage on the pacemaker device can be greatly simplified. In addition, this method clearly explains the induced lead heating mechanism during MRI procedures. Several numerical examples, as well as measurement results are given to demonstrate the efficiency and accuracy of this method.

细分领域

< | 射频致热 | 射频致热评估